Command Reference
R&S
®
CMW-B400B/-B405A
75
User Manual 1176.8116.02 ─ 04
Parameters:
<StatisticCount>
Range:
1 to 1000
*RST:
20
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:MULTitone:RMODe
<ResultMode>
CONFigure:AUDio:MEAS<i>:DIGital:MULTitone:RMODe
<ResultMode>
Selects whether the multitone results are provided as absolute levels (dBV or dBFS) or
as relative levels (dB).
Parameters:
<ResultMode>
ABS | T01 | T02 | T03 | T04 | T05 | T06 | T07 | T08 | T09 | T10 |
T11 | T12 | T13 | T14 | T15 | T16 | T17 | T18 | T19 | T20
ABS:
absolute levels
T01:
levels relative to the level of test tone 1
...
T20:
levels relative to the level of test tone 20
*RST:
ABS
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:MULTitone:FILTer:LPASs
<Frequency>
CONFigure:AUDio:MEAS<i>:DIGital:MULTitone:FILTer:LPASs
<Frequency>
Configures the cutoff frequency of the lowpass filter for multitone measurements.
Parameters:
<Frequency>
OFF | LP3 | LP4 | LP15
OFF:
lowpass filter disabled
LP3 | LP4 | LP15:
cutoff frequency 3 kHz, 4 kHz or 15 kHz
*RST:
OFF
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:MULTitone:FILTer:HPASs
<Frequency>
CONFigure:AUDio:MEAS<i>:DIGital:MULTitone:FILTer:HPASs
<Frequency>
Configures the cutoff frequency of the highpass filter for multitone measurements.