GUI Reference
R&S
®
CMW-B400B/-B405A
31
User Manual 1176.8116.02 ─ 04
Fig. 4-5: General analog measurement settings
................................................................................................. 32
....................................................................... 32
Coupled with
Displays the coupling of the measurement to a generator. The measurement can be
used standalone ("Off") or it can be coupled to an analog or digital generator.
The coupling is configured within the generator settings, see
Remote command:
n/a
Routing
Displays the input connector for the measured analog audio signal (AF 1 IN or AF 2
IN). This setting is not configurable.
Remote command:
n/a
Repetition
Defines how often the measurement is repeated if it is not stopped explicitly or by a
failed limit check.
●
Continuous:
The measurement is continued until it is explicitly terminated; the
results are periodically updated.
●
Single-Shot:
The measurement is stopped after one statistics cycle.
Audio Measurement Settings