GUI Reference
R&S
®
CMW-B400B/-B405A
33
User Manual 1176.8116.02 ─ 04
●
Auto ranging:
If auto ranging is enabled, the maximum expected AF input level is set automati-
cally according to the average level of the measured AF signal plus an appropriate
overload margin.
●
Manual configuration:
If auto ranging is disabled, the manually configured "Maximum Level" is applied.
To optimize the "Maximum Level" setting, start with a very low value. Then
increase it, until the error message "Input Overdriven" disappears.
Remote command:
CONFigure:AUDio:MEAS<i>:ANALog:ARANging
CONFigure:AUDio:MEAS<i>:ANALog:MLEVel
4.3.2 General Settings (Digital)
The settings at the top of the "Digital Meas" tab apply to all digital measurements.
Fig. 4-6: General digital measurement settings
Coupled with
Displays the coupling of the measurement to a generator. The measurement can be
used standalone ("Off") or it can be coupled to an analog or digital generator.
The coupling is configured within the generator settings, see
Remote command:
n/a
Routing
Displays the input connector for the measured digital audio signal (SPDIF IN) and the
evaluated channel (-L or -R). This setting is not configurable.
Remote command:
n/a
Audio Measurement Settings