Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800E Series
2-49
Meas Setup
Open the parameter setting menu of the function selected in the
Meas
menu. This
key is only available when a measurement function is enabled.
VSWR
Measurement Interface:
Figure 2-12 VSWR Measurement Interface
Measurerment Results:
return loss, reflection coefficient and VSWR.
Return Loss: the formula is
)
lg(
10
)
(
r
i
P
P
dB
RL
=
. Wherein,
)
(
dB
RL
is the
return loss,
i
P
is the incident power and
r
P
is the reflected power.
Reflection Coefficient: the ratio of the reflected voltage to the incident
voltage.
VSWR: the ratio of the maximum voltage to the minimum voltage of the
standing wave.
This measurement requires the VSWR, VSWR bridge options and TG
(DSA832E-TG). Two measurements have to be performed according to the