Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800E Series
2-91
Table 2-56 PK Excursn
Parameter
Explanation
Default
10 dB
Range
0 dB to 200 dB
Unit
dB
Knob Step
1 dB
Arrow Key Step
1 dB
2
.
PK Thresh
Assign a minimum for the peak amplitude. Peaks whose amplitudes are
greater than the specified peak threshold are treated as real peaks.
Table 2-57 PK Thresh
Parameter
Explanation
Default
-90 dBm
Range
-200 dBm to 0 dBm
Unit
dBm, -dBm, mV, uV
Knob Step
1 dBm
Arrow Key Step
1 dBm
3
.
Peak Search
Set the peak search condition. The available options are Maximum and Para.
If
Max
is selected, the system searches for the maximum on the trace.
If
Para
is selected, the system searches for the peak that meets the
parameter condition on the trace.
Note:
This setting applies only to the peak search executed by pressing
Peak
at the front panel; while other searches such as Next Peak, Peak Right, Peak
Left and Min Search are all based on the
Para
.
Peak Table
Open the peak table (in the lower window) which lists the peaks (with frequency
and amplitude) that meet the peak search condition. Up to 10 peaks can be
displayed in the table.