RIGOL
Chapter 2 Front Panel Operation
User’s Guide for DSA800E Series
2-58
Chan Pwr
Measurement Interface:
Figure 2-15 Channel Power Measurement Interface
Measurement Results:
channel power and power spectral density.
Channel Power: power within the integration bandwidth.
Power Spectral Density: power (in dBm/Hz) normalized to 1 Hz within the
integration bandwidth.
Measurement Parameters:
average number, average mode, integration
bandwidth and channel power span.
1.
Avg Num
Specify the number of averages used when calculating the measurement
result and the default is "Off". You can use the numeric keys, the knob, or the
arrow keys to modify the parameter.
Integration Bandwidth Channel Power Span
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30