Chapter 2 Front Panel Operation
User’s Guide for DSA800E Series
2-56
Table 2-31 Average Number for ACP Measurement
Parameter
Description
Default
10
Range
1 to 1000
Unit
N/A
Knob Step
1
Arrow Key Step
1
2.
Avg Mode
Set the average operation mode to "Exp" or "Repeat" and the default is "Exp".
When "Exp" is selected, the result is the exponential average of the
current
N
(
N
is specified in "
") measurement results.
When "Repeat" is selected, the result is the arithmetic average of the
current
N
(
N
is specified in "
") measurement results.
3.
Main CH BW
Set the bandwidth of the main channel and the power of the main channel is
the power integral within this bandwidth. You can use the numeric keys, the
knob, or the arrow keys to modify the parameter.
Table 2-32 Main Channel Bandwidth for ACP Measurement
Parameter
Description
Default
2 MHz
Range
33 Hz to 2.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Main Channel BW/100, the
minimum is 1Hz
Arrow Key Step
in 1-1.5-2-3-5-7.5 sequence
4.
Adj CH BW
Set the frequency width of the adjacent channels.
The adjacent channel bandwidth is related to the main channel
bandwidth and the range available is from mian channel bandwidth/20 to
mian channel bandwidth×20.
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter.