RIGOL
User’s Guide for DSA800E Series
5-3
SSB Phase Noise
20
℃
to 30
℃
, f
c
= 1 GHz
Carrier offset
10 kHz offset
<-90 dBc/Hz
Residual FM
20
℃
to 30
℃
, RBW = VBW = 1 kHz
Residual FM
<20 Hz (nom.)
Bandwidths
Set "Auto SWT" to "Accy"
Resolution bandwidth (-3
dB)
10 Hz to 1 MHz, in 1-3-10 sequence
RBW uncertainty
<5% (nom.)
Resolution filter shape
factor (60 dB : 3 dB)
<5 (nom.)
Video bandwidth
(-3 dB)
1 Hz to 3 MHz, in 1-3-10 sequence
Resolution bandwidth (-6
dB) (EMI-DSA800 option)
200 Hz, 9 kHz, 120 kHz
Amplitude
Measurement Range
Range
f
c
≥ 10 MHz
DANL to +20 dBm
Maximum Input Level
DC voltage
50 V
CW RF power
attenuation = 30 dB
+20 dBm (100 mW)
Max. damage level*
+30 dBm (1 W)
Note: *When f
c
≥ 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on.
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30