Marker->
Set the other system parameters (such as center frequency and reference level)
using the current marker readings. Press
Marker->
to enable a marker
automatically if no marker is currently enabled.
Mkr->CF
Set the center frequency of the analyzer to the frequency of the current marker.
If
Normal
marker is selected, the center frequency will be set to the
frequency of the current marker.
If
Delta
,
Delta Pair
,
or
Span Pair
marker is selected, the center frequency
will be set to the frequency of the Delta Marker.
The function is invalid in Zero span mode.
Mkr->Step
Set the center frequency step of the analyzer to the frequency of the current
marker.
If
Normal
marker is selected, the center frequency step will be set to the
frequency of the current marker.
If
Delta
,
Delta Pair
or
Span Pair
marker is selected, the center frequency
step will be set to the frequency of the Delta Marker.
The function is invalid in Zero span mode.
Mkr->Start
Set the start frequency of the analyzer to the frequency of the current marker.
If
Normal
marker is selected, the start frequency will be set to the frequency
of the current marker.
If
Delta
,
Delta
Pair
or
Span Pair
marker is selected, the start frequency will
be set to the frequency of the Delta Marker.
The function is invalid in Zero span mode.