Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800E Series
2-29
Auto SWT
Set the auto setting method of the sweep time to "Normal" or "Accy". The sweep
will be faster if "Normal" is selected; while in "Accy" type, higher measurement
precision can be obtained.
Mode
Set the sweep mode to "Single" or "Cont". The default is "Cont". The
corresponding icon of the sweep mode will be displayed in the status bar at the left
side of the screen.
Single Continuous
1.
Single
Set the sweep mode to "Single". The number
10
on the parameter icon
denotes the current sweep number.
2.
Cont
Set the sweep mode to "Cont". The character
Cont
on the parameter icon
denotes the analyzer is sweeping continuously.
Key Points:
If the instrument is in single sweep mode and no measurement function is
enabled, press this key and the system will enter continuous sweep mode and
sweep continuously if the trigger conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on,
press this key and the system will enter continuous sweep mode and measure
continuously if the trigger conditions are satisfied.
In continuous sweep mode, the system will send a trigger initialization signal
automatically and enter the trigger condition judgment directly after each
sweep.