Chapter 2 Front Panel Operation
User’s Guide for DSA800E Series
2-74
Table 2-51 Average Number for TOI Measurement
Parameter
Description
Default
10
Range
1 to 1000
Unit
N/A
Knob Step
1
Arrow Key Step
1
2.
Avg Mode
Set the average operation mode to "Exp" or "Repeat" and the default is "Exp".
When "Exp" is selected, the result is the exponential average of the
current
N
(
N
is specified in "
") measurement results.
When "Repeat" is selected, the result is the arithmetic average of the
current
N
(
N
is specified in "
") measurement results.
3.
Span
This span which is the same with the span of the analyzer is the frequency
range for sweep. Modifying this parameter will also change the span of the
analyzer. You can use the numeric keys, the knob, or the arrow keys to modify
the parameter.
Table 2-52 Span for TOI Measurement
Parameter
Description
Default
2 MHz
Range
100 Hz to 3.2 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
TOI Distortion Span/100, the
minimum is 1 Hz
Arrow Key Step
in 1-1.5-2-3-5-7.5 sequence