RIGOL
Chapter 2 Front Panel Operation
User’s Guide for DSA800E Series
2-14
Table 2-7 Reference Level
Parameter
Explanation
Default
0 dBm
Range
-100 dBm to 20 dBm
Unit
dBm, -dBm, mV, uV
Knob Step
In Log scale mode, step = Scale/10
In Lin scale mode, step = 0.1 dBm
Arrow Key Step
In Log scale mode, step = Scale
In Lin scale mode, step = 1 dBm
Input Atten
Set the front attenuator of the RF input in order to ensure big signals (or small
signals) to pass through the mixer with low distortion (or low noise).
Key Points:
When the preamplifier is turned on, the input attenuation could be set up to
30 dB. You can adjust the reference level to ensure that the specified
parameter meets the inequality (2-3).
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter. For more details, please refer to "
".
Table 2-8 Input Attenuation
Parameter
Explanation
Default
10 dB
Range
0 dB to 30 dB
Unit
dB
Knob Step
1 dB
Arrow Key Step
5 dB
Scale/Div
Set the logarithmic units per vertical grid division on the display. This function is
only available when the scale type is set to "Log".
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30