Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800E Series
2-43
3.
Norm Ref Lvl
Adjust the vertical position of the trace on the screen by adjusting the
reference level when normalization is enabled.
Being different from the
Ref Level
function in the
AMPT
menu, this
parameter has no influence on the reference level of the analyzer.
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter. For more details, please refer to "
".
Table 2-25 Reference level under normalization
Parameter
Explanation
Default
0 dB
Range
-200 dB to 200 dB
Unit
dB
Knob Step
1 dB
Arrow Key Step
10 dB
4.
Norm Ref Pos
Adjust the vertical position of the normalization reference level on the screen
by adjusting the reference position when normalization is enabled.
The function of this menu is similar to that of
Norm Ref Lvl
. When it is
set to 0%, the normalization reference level is displayed at the bottom of
the screen grid and at the top when it is set to 100%.
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter. For more details, please refer to "
".
Table 2-26 TG reference position
Parameter
Explanation
Default
100%
Range
0% to 100%
Unit
%
Knob Step
1%
Arrow Key Step
10%
5.
Ref Trace
Set whether to display the reference trace or not. If "View" is selected, the
reference trace saved (Trace 3) will be shown in "Freeze" type.