RIGOL
Chapter 2 Front Panel Operation
User’s Guide for DSA800E Series
2-42
value.
Key Points:
This parameter only changes the readout of the TG output power, rather than
the actual value of it.
The offset could be either a positive (gain in the external output) or a negative
(loss in the external output).
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter. For more details, please refer to "
".
Table 2-24 TG Level Offset
Parameter
Explanation
Default
0 dB
Range
-200 dB to 200 dB
Unit
dB
Knob Step
1 dB
Arrow Key Step
10 dB
Normalize
Nomalization can eliminate the error of TG Level. Before using this function,
connect the
[GEN OUTPUT 50Ω]
output terminal of the TG with the
[RF INPUT
50Ω]
input terminal of the analyzer.
1.
Stor Ref
Save the data of Trace 1 to Trace 3 as the reference value for normalization.
This operation should be done before you enable the normalization.
2.
Normalize
Enable or disable the normalization. When enabled, the reference trace will be
stored automatically after the current sweep finishes if no reference trace is
stored before. During the reference trace storage, the corresponding prompt
message is displayed. When normalization is enabled, the corresponding
value of the reference trace will be substracted from the trace data after every
sweep.