Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800E Series
2-15
Key Points:
By changing the scale, the amplitude range available is adjusted.
The range of the amplitude that can be displayed:
Minimum: reference level – 10 × the current scale value
Maximum: the reference level.
You can use the numeric keys, the knob, or the arrow keys to modify the
parameter. For more details, please refer to "
".
Table 2-9 Scale
Parameter
Explanation
Default
10 dB
Range
0.1 dB to 20 dB
Unit
dB
Knob Step
Scale ≥ 1, step = 1 dB
Scale < 1, step = 0.1 dB
Arrow Key Step
in 1, 2, 5 sequence
Scale Type
Set the scale type of Y-axis to Lin or Log. The default is Log.
Key Points:
In Log scale type, the Y-axis denotes the logarithmic coordinate; the value
shown at the top of the grid is the reference level and each grid represents the
scale value. The unit of Y-axis will automatically switch to the default unit
(dBm) in Log scale type when the scale type is changed from Lin to Log.
In Lin scale type, the Y-axis denotes the linear coordinate; the values shown
at the top of the grid and the bottom of the grid are the reference level and 0
V respectively. Each grid represents 10% of the reference level and the scale
setting function is invalid. The unit of Y-axis will automatically switch to the
default unit (Volts) in Lin scale type when the scale type is changed from Log
to Lin.
The scale type does not affect the unit of Y-axis.