Chapter 3 Arbitrary Waveform Output
DG4000 User's Guide
3-12
5.
Interp
Press this softkey to enable or disable the interpolation between the defined
waveform points.
Off: The waveform editor will hold a constant voltage level between two
points and create a step waveform.
Linear: The waveform editor will automatically connect the two defined
points with a straight line.
6.
Edit Points
Define a waveform by specifying the time and voltage of each waveform point.
Press this softkey to enter the "Edit Points" interface.
Point: Set the ID of the point to be edited. The default is 1. You can use
the numeric keyboard or direction keys and knob to modify this
parameter.
Time: Set the time of the current point in a period. This setting is limited
by the times of the previous and the next points. Note that the time of
point 1 is fixed at 0 s.
Voltage: Set the voltage of the current point in mV or V. The range of the
voltage is limited by the "
HiLevel" and "LoLevel".
Insert: Insert a waveform point at the middle position of the currently
edited point and the next edited point. When the interpolation mode is set
to "Off", the voltage of the waveform point inserted is the same with that
of the current point. When the interpolation mode is set to "Linear", the
voltage of the waveform point inserted is the middle value of the voltage
values of the current edited point and the next edited point.
Delete: Delete the current point from the waveform and connect the
remaining points using the current interpolation method. Note that the
first waveform point and the last waveform point cannot be deleted.
7.
Edit Block
Define a waveform by editing the time and voltage of the start and end points
and automatically setting the points between them through linear interpolation.
Press
Interp to select "Linear", and then press this softkey to enter the "Edit
Block" interface. Note that if "Linear" is not selected in
Interp, this menu is
grayed out and disabled.
X1: Set the ID of the start point of the block (namely the time position of
the start point). X1 should be lower than or equal to the number of initial
Summary of Contents for DG4062