Chapter 2 Basic Waveform Output
DG4000 User's Guide
2-8
To Set the DC Offset Voltage
The settable range of the DC offset voltage is limited by the "Resistance" and
"Ampl/HiLevel" settings. Please refer to the "Output Characteristics" in
"
Specifications". The default value is 0V
DC
.
The DC offset voltage displayed on the screen is the default value or the offset
previously set. When the instrument configuration (such as resistance) is changed,
if this offset is valid, the instrument will still use this offset; otherwise, the
instrument would display prompt message and set the offset to the offset upper
limit of the new configuration automatically.
Press
Offset/LoLevel to highlight "Offset". At this point, use the numeric
keyboard to input the offset value and select the desired unit from the pop-up unit
menu or use the direction keys and knob to modify the current value.
For the input method of offset value, refer to the introduction in "
The DC offset voltage units available are V
DC
and mV
DC
.
Press this softkey again to switch to low level setting. At this point, "LoLevel"
is highlighted.
The low level units available are V and mV.