DG4000 User's Guide
13-8
Pulse Width and Duty Cycle Measurement
Frequency/Amplitud
e Range
1μHz to 25MHz
50mVRMS to
±2.5Vac+dc
DC Coupling
Input
Attenuation:
"closed"
Pulse Width
Minimum
≥20ns
Resolution
2ns
Duty Cycle
Range (Display)
0% to 100%
Input Characteristics
Input Range
Breakdown Voltage
±7Vac+dc
(Attenuation:
closed)
Impedance=1
MΩ
±70Vac+dc
(Attenuation:
open)
5Vrms
Impedance=5
0Ω
Input Adjustment
Attenuation
Open: "×10"; Closed: "×1"
Impedance
50Ω
1MΩ
Coupling
AC
DC
HF Reject
ON: input bandwidth=250kHz;
OFF: input bandwidth=225MHz
Input Trigger
Trigger Level Range
-2.5V to +2.5V
Trigger Sensitivity Range
0% (140mV hysteresis voltage) to
100% (2mV hysteresis voltage)
Gate Time
GateTime1
1ms
GateTime2
10ms
GateTime3
100ms
GateTime4
1s
GateTime5
10s
GateTime6
>10s
Trigger Characteristics
Trigger Input
Level
TTL-compatible