34
iDimension 300
Communication Parameters
Any required parameters needed to control communication with the scale can be entered into this field.
For serial scale connections (namely Mettler Toledo, MTSICS, NCI and Pennsylvania 7300), other than the default
values (which can be left empty), the following format is expected:
<BAUD RATE>,<PARITY>,<BITS>,<STOPBITS>
Eg. 9600,N,8,1
Table 5-4. Valid Values for Scale Communication Parameters
For USBHID, other than the supported scales listed, the scale communication parameters should contain the
vendor and product ID in the following format:
<VENDOR ID>,<PRODUCT ID>
Eg: 0x0EB8,0XF000
Scale Stable Status
Determines whether
iDimension 300
will wait for the scale to stabilize before accepting a weight or status.
Default Value: On
Valid Values: On/Off
Wait Timeout (ms)
The period of time (in milliseconds) that the system will wait for a response from the scale.
Default Value: 3000
Valid Values: Scale Dependent
Low Resolution Camera
The Switch Resolution Delay (ms) is a minimum period of time
(in milliseconds)
that the system should
pause after switching the resolution of the camera. Do not change this setting unless instructed to by Rice
Lake Weighing Systems.
Default Value: 200
Valid Values: 100ms-500ms
High Resolution Camera
Bar Code Recognition
Set the maximum number of bar codes the system should consider when processing a single capture.
Default Value: 100
Valid Values: 1-999
Parameters
Values
Baud Rate
1200, 2400, 4800, 9600, 14400, 19200, 38400, 57600 and 115200
Parity
N,O,E
Bits
5,6,7,8,9
Stopbits
1,1.5,2
Setting
Description
Focal Length (Pixels)
Focal length (in pixels) of the high resolution camera when the lens is fully zoomed out and the camera is
set to 640 x 480 pixels. Do not change this setting unless instructed to by Rice Lake Weighing Systems.
Default Value: 980
Focus Offset
An offset that will be applied to each value in the focus table. The use of this parameter is reserved.
Table 5-5. High Resolution Camera Settings
Summary of Contents for IDIMENSION 200
Page 2: ...iDimension 300 Industrial Installation Manual PN 169040...
Page 3: ......
Page 15: ...12 iDimension 300 6 Click Done to save the settings...
Page 35: ...32 iDimension 300 General Settings Figure 5 10 iDimension 300 General Settings Main Menu...
Page 69: ...66 iDimension 300...
Page 70: ...Administrator s Section 67...
Page 71: ......