16
iDimension 300
Figure 3-7. Start the Calibration Process
4. Select the
High Resolution
tab.
5. Ensure that the yellow diagonal lines intersect as near as possible to the center of the calibration object. If
the lines are not intersecting at the center, move the calibration object until the center mark intersects with
the yellow diagonal lines. If the marks cannot be centered, try to adjust the physical position of the head by
adjusting the top section of the head and/or adjusting the screws on the rear of the head.
6. Hold down the left mouse button and drag the mouse to draw a rectangle around the calibration object.
7. Select the
Depth Confidence
tab.
8. Hold down the left mouse button and drag the mouse to draw a rectangle around the calibration object.
Figure 3-8. Draw a Rectangle around the Calibration Object
9. Select the
Low Resolution
tab.