Key invalid
The key you have pressed is not valid during
this operation
Recheck the instructions for this operation to see
what the valid key functions are (see Section 1.2.2
Key Functions)
Number invalid
You have exceeded the maximum number of
characters allowed for this field
Abbreviate the text you are entering so that it fits.
See Section 7.1.3, in the DC-200/300 Operation
Manual for character lengths allowed for different
fields .
Exceed max limit
You have exceeded the maximum number of
characters allowed for this fields
Abbreviate the text you are entering so that it fits.
See Section 7.1.3, in the DC-200/300 Operation
Manual for character lengths allowed for different
fields .
Please press code key
Save the data and exit programming by
pressing the Code key
Date invalid
The date you have entered is not in the
necessary format
Enter date in correct format
Note: SPEC 05 sets date format. Default is MM:
DD: YY. See Section 2.9, Configuration Settings.
Time invalid
The time you entered is not in the necessary
format
Enter time in correct format
Note: SPEC 06 sets time format. Default is HH:
MM in 12-hour format. See Section 2.9,
Configuration Settings.
Logo not exist
(DC-300 only)
The requested logo is not programmed into
the scale
Program a logo into the scale (see Section 7.2.4,
Logo Files in the DC-200/300 Operation Manual)
Label invalid
The data requested to print on the label does
not exist
Check the item to see if this data has been entered
(see Section 6.3.7, View Item Information in the
DC-200/300 Operation Manual)
Data invalid
The data keyed in is not in a valid format
Recheck the acceptable format for the data item
(see Sectin 6.1.2, Programing Item Code in the
DC-200/300 Operation Manual)
Quantity overflow
The quantity data is over the limit for the
number of places that can be displayed
Work with a smaller quantity. The maximum
number of digits for the quantity display is 6.
Record chosen in use
The record you want to delete is in use by
another data file
Close the operation that is using the record and
restart the delete procedure.
Address not exist
(DC-300 only)
The address does not exist in the scale’s
database
Program an address into the scale (see Section
7.2.2, How to Program an Address File in the
DC-200/300 Operation Manual)
Text not exist
(DC-300 only)
The text field does not exist in the scale’s
database
Program a text field into the scale (see Section
7.2.3, How to Program a Text File in the DC-200/
300 Operation Manual)
No print area
(DC-300 only)
This label format does not include a print
area for this field
Change to a different pre-programmed label format
(see Section 7.1.2, Label Format Samples), or
program a free format label that includes the item
(see Section 7.5, Free Format Label Programming
in the DC-200/300 Operation Manual)
Print inhibited
A weigh item cannot be printed during a
counting operation
Exit the operation you are in, then re-try printing
the label.
Force balance error
There is no force balance connected to the
scale
Connect a force balance to the scale (see Section
7.3, Connection From Scale to Force Balance)
Scale A start up err press
clr key to exit
Unable to detect scale A because it is not
stable or the weight on the platform is out of
the scale start-up range. Press Clear key to
exit.
Bring the weight on the platform down to within the
start-up range.
Error Message
Causes
Corrective Action
Table 9-1. DC-200/300 Error Message List