NOTE
The error conditions are categorized into three groups:
CONTINUE/LOGGED:
This group contains error conditions in which there is
reasonable expectation that the device will operate within published specification
and not cause an unacceptable level of risk.
REBOOT:
This group contains error conditions in which there is a reasonable
expectation that the error was a transient and the device will perform properly after
a system reboot.
STOP:
This group contains error conditions in which there is a low confidence in
the ability of the device to operate within specification or may cause unacceptable
level of risk. Additionally, any three (3) Reboot errors within a 24-hour period as
referenced from the last error shall constitute a Stop error.
User will only see STOP errors displayed.