Procedure
Removed/Installed During Process:
•
Top Enclosure (Refer to page 5 for more detailed instructions.)
•
Main PCA (Refer to page 7 for more detailed instructions.)
•
Outlet-Flow-Tubing (Refer to page 12 for more detailed instructions.)
•
Buzzer Assembly
F
IGURE
S: R
EMOVAL
AND
I
NSTALLATION
OF
THE
B
UZZER
A
SSEMBLY
To remove the Buzzer Assembly:
1. Remove the Top Enclosure (Refer to page 5 for more detailed instructions).
2. Remove the Main PCA (Refer to page 7 for more detailed instructions).
3. Remove the Outlet-Flow-Tubing Assembly (Refer to page 12 for more detailed instruc-
tions).
4. Remove the two screws and washer that fasten the buzzer assembly to the Bottom Enclo-
sure and lift the buzzer up and away from the Bottom Enclosure.
To install the Buzzer Assembly:
1. Place the buzzer in the Bottom Enclosure and fasten it down using two screws and wash-
ers. Ensure that the buzzer wire is positioned towards the rear of the unit.
2. Install the Outlet-Flow Tubing Assembly (Refer to page 12 for more detailed instructions).
3. Install the Main PCA (Refer to page 7 for more detailed instructions).