B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1032411, R
EV
. 00
S
ETUP
- P
AGE
4
2. If CPAP is chosen, the CPAP setting screen will appear next.
Increase or decrease the CPAP pressure by pressing the Heat and Ramp buttons until the
desired pressure appears. You can adjust the pressure from 4 to 20 cm H
2
O in 1 cm H
2
O
increments. Press the right user button to move to the next screen. If CPAP is the chosen
mode, proceed to Step 10 for next screen.
If PC, S, S/T, or T mode is chosen, the AVAPS screen will appear next.
Turn the AVAPS on or off by pressing the Heat or Ramp buttons until the correct number
appears. A One (1) on the display screen indicates that AVAPS is enabled and a Zero (0)
indicates that the AVAPS feature is disabled. Press the right user button to move to the
next screen. If AVAPS is not chosen proceed to Step 6.
3. The Target Tidal Volume screen will appear next.
Increase or decrease the Target VT pressure by pressing the Heat or Ramp buttons until
the correct pressure appears. You can adjust Target VT from 200 mL to 1500mL in
increments of 10mL. Press the right user button to move to the next screen.
EPAP
cm
H
2
O
CPAP
PC
AVAPS
ml
PC
V
T