10. When in S mode the next screen will be the Flex Setting screen. (All other modes should
proceed to Step 11.)
This screen allows you to adjust the level of air pressure relief that the patient feels when
he or she exhales during therapy. To change the selection, press the Heat or Ramp
buttons until the correct setting appears. The Flex set point is adjustable from 0-3 in
increments of 1. Press the right user button to move to the next screen.
11. The next screen is the Ramp Length Setting. This allows you to change the ramp time.
To change the ramp time, press the Heat and Ramp buttons until the correct time appears.
The setting increases or decreases from 0 to 45 minutes in 5 minute increments. If you do
not want to ramp, set the time to zero. Press the right user button to move to the next
screen.
12. The next screen is the ramp setting pressure. This screen will only display if the ramp
length setting is greater than zero.
To change the ramp setting pressure, press the Heat and Ramp buttons until the correct
pressure appears. The setting increases or decreases in 1.0 cmH
2
O increments. The
user can adjust the setting from 4 cmH
2
O to the EPAP/CPAP pressure setting. Press the
right user button to move to the next screen.
NOTE
The Flex Setting is not available if the IPAP at 26 cm H
2
O or higher.
FLEX
S
PC
PC
cm
H
2
O