B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
S
ETUP
- P
AGE
5
1032411, R
EV
. 00
4. The IPAP max screen will appear next.
Increase or decrease the IPAP max pressure by pressing the Heat or Ramp buttons until
the correct pressure appears. You can adjust the pressure from 4 to 30 cm H
2
O in 1 cm
H
2
O increments. The IPAP max pressure setpoint cannot be set lower than the IPAP min
pressure setting. Press the right user button to move to the next screen.
5. The IPAP min screen will appear next.
Increase or decrease the IPAP min pressure by pressing the Heat or Ramp buttons until
the correct pressure appears. You can adjust the pressure from 4 to 30 cm H
2
O in 1 cm
H
2
O increments. The IPAP min pressure setpoint cannot be set lower than the EPAP
setting. Press the right user button to move to the next screen.
6. If AVAPS is not enabled the IPAP screen will appear next.
Increase or decrease the IPAP pressure by pressing the Heat or Ramp buttons until the
correct pressure appears. You can adjust the pressure from 4 to 30 cm H
2
O in 1 cm H
2
O
increments. The IPAP pressure setpoint cannot be set lower than the EPAP setting.
Press the right user button to move to the next screen.
cm
H
2
O
PC
IPAP
cm
H
2
O
PC
IPAP
cm
H
2
O
PC
IPAP