B
I
PAP
S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1032444, R
EV
. 00
R
EPAIR
& R
EPLACEMENT
- P
AGE
22
B
OTTOM
E
NCLOSURE
Procedure
Removed/Installed During Process:
•
Top Enclosure (Refer topage 5 for more detailed instructions.)
•
Main PCA (Refer to page 7 for more detailed instructions.)
•
Outlet-Flow-Tubing (Refer topage 12 for more detailed instructions.)
•
Diffuser Assembly (Refer to page 14 for more detailed instructions.)
•
Valve Assembly (Refer to page 15 for more detailed instructions.)
•
Blower Assembly (Refer to page 17 for more detailed instructions.)
•
Muffler (Refer to page 19 for more detailed instructions.)
•
Buzzer (Refer to page 21 for more detailed instructions.)
•
Bottom Enclosure
F
IGURE
T: B
OTTOM
E
NCLOSURE
To remove the Bottom Enclosure:
1. Remove the Top Enclosure (Refer to page 5 for more detailed instructions).
2. Remove the Main PCA (Refer to page 7 for more detailed instructions).
3. Remove the Outlet-Flow-Tubing Assembly (Refer to page 12 for more detailed instruc-
tions).
4. Remove the Diffuser Assembly (Refer to page 14 for more detailed instructions).
5. Remove the Valve Assembly (Refer to page 15 for more detailed instructions).
6. Remove the Blower Assembly (Refer to page 17 for more detailed instructions).
7. Remove the Muffler (Refer to page 19 for more detailed instructions).
8. Remove the Buzzer (Refer to page 21 for more detailed instructions).
9. Remove Light Pipe from Bottom Enclosure.