B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
S
PECIFICATIONS
& S
YSTEM
F
EATURES
- P
AGE
3
1032409, R
EV
. 00
F
UNCTIONS
Modes
Continuous Positive Airway Pressure (CPAP)
Spontaneous (S)
Timed (T)
Spontaneous/Times (S/T)
Pressure Control (PC)
P
RESSURE
IPAP
4 to 30 cm H
2
O*
EPAP
4 to 25 cm H
2
O
CPAP
4 to 20 cm H
2
O
V
te
Setpoint
200mL to 1500mL
Low MinVent Alarm
0.0 LPM to 99.0 LPM
Flex Setpoint
0 to 3
Breath Rate
0 to 30 BPM
Timed Inspiration
0.5 to 3.0 seconds
Ramp Duration
0 to 45 minutes
Rise Time**
1 to 6 (0.1 to 0.6 seconds)
* Limited to 25 cmH
2
O when using the Bi-Flex feature in S mode.
** The range of values correspond to tenths of seconds (e.g., a setting of 4
indicates a Rise Time of 0.4 seconds). When in BiFlex Mode, the unit shall always
use a rise time of 3 regardless of this setpoint.
P
HYSICAL
Dimensions
9.8”(L) x 6.6”(W) x 4.4”(H)
Weight
Less than 7 lbs. without the humidifier
EMC N
ORMATIVE
D
OCUMENTS
This product is designed to conform to the following standards:
IEC 60601-1
Medical Electrical Equipment, Part 1
General Requirements for Safety and
essential performance.