B
I
PAP
S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1032444, R
EV
. 00
R
EPAIR
& R
EPLACEMENT
- P
AGE
14
D
IFFUSER
A
SSEMBLY
R
EPLACEMENT
Procedure
Removed/Installed During Process:
•
Top Enclosure (Refer to page 5 for more detailed instructions.)
•
Main PCA (Refer to page 7 for more detailed instructions.)
•
Outlet-Flow-Tubing (Refer to page 12 for more detailed instructions.)
•
Diffuser Assembly
F
IGURE
M: D
IFFUSER
A
SSEMBLY
To remove the Diffuser Assembly:
1. Remove the Top Enclosure (Refer to page 5 for more detailed instructions on removing
the Top Enclosure).
2. Remove the Main PCA (Refer to page 7 for more detailed instructions on removing the
Main PCA).
3. Remove the Outlet-Flow Tubing (Refer to page 12 for more detailed instructions on
removing the Outlet-Flow-Tubing).
4. Slide the Coupler off the Diffuser Assembly.
5. Slide the Diffuser out of the Valve Assembly.
To install the Diffuser Assembly:
1. Slide the Diffuser into the Valve Assembly
2. Slide the Coupler onto the other end of the Diffuser.
3. Install the Outlet-Flow-Tubing (Refer to page 12 for more detailed instructions on installing
the Outlet-Flow-Tubing).
4. Install the Main PCA (Refer to page 7 for more detailed instructions on installing the Main
PCA).
5. Install the Top Enclosure (Refer to page 5 for more detailed instructions on installing the
Top Enclosure).
Diffuser