B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1032443, R
EV
. 00
T
ROUBLESHOOTING
& D
IAGNOSTICS
- P
AGE
18
Error Code
Description of Error
Device Action
E54
Motor Current High While Blower Off
STOP
E55
Buffer Overflow
REBOOT
E56
Motor Current High While Blower On
STOP
E57
Unable to Queue Data to Provide Therapy
REBOOT
E58
Stuck Key
CONTINUE
E59
Outlet Pressure Sensor Railed
STOP
E60
Blower Pressure Sensor Railed
STOP
E61
Real-time Clock’s Battery Dead
CONTINUE
E62
Blower Speed Exceeds Maximum Limit
STOP
E63
Internal Watchdog Failure
STOP
E64
External Watchdog failure
STOP
E65
Unexpected Watchdog Reset
REBOOT
E66
Reserved for Future Use
REBOOT
E67
Operating System Failed Initialization
REBOOT
E68
Unable to Queue Data for Communication
REBOOT
E69
Sampling Thread Locked
REBOOT
E70
Execution Thread Locked
REBOOT
E71
Internal RAM Failure
STOP
E72
Unable to Queue Data for logging
REBOOT
E73
Reserved for Future Use
REBOOT
E74
Reserved for Future Use
REBOOT
E75
Incorrect Display Tape
STOP
E76
Failure in Loss of Power Battery
STOP
E77
High Pressure Condition
LOGGED
E78
Low Pressure Condition
LOGGED
E79
Unable to Maintain Pressure Support
LOGGED
E80
Barometric Pressure Sensor Failure
LOGGED
E81
Barometric Sensor Failure
STOP
E82
Invalid Error Code
RESERVED