B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1032447, R
EV
. 00
T
ESTING
& C
ALIBRATION
- P
AGE
18
33. After zero flow is confirmed, the program will verify negative flow. Connect the patient tub-
ing between the outlet of the flow valve and the negative flow source. Turn on the nega-
tive flow source and follow the prompts.
34. After completion of the negative flow verification, pressure accuracy will be checked. Turn
off the negative flow source. Disconnect the patient tubing and occlude the O
2
port at the
BiPAP Synchrony outlet port.
35. The program will now verify the barometirc pressure. Enter the value, in inches of mer-
cury, displayed on the barometer.
36. When the pressure verification is complete, the program will request a button test for the
BiPAP Synchrony. Press the buttons on the device in the order that they are numbered on
the program screen.
F
IGURE
AA: B
UTTON
T
EST
37. When the button test has been completed and the BiPAP Synchrony has successfully
passed all tests, a green “PASS” prompt will appear on the computer screen. Click the
“OK” button to view the test data sheet. Click “OK” on the print results button to print the
test data sheet.
F
IGURE
AB: P
ASS
S
CREEN