Error Code
Description of Error
Device Action
E25
Excessive Drift on Flow Sensor
STOP
E26
Communications Failure with LCD Driver Chip
STOP
E27
Excessive Drift on Outlet Pressure Sensor
STOP
E28
Empty Calibration Table
Not Used
E29
Excessive Drift on Blower Pressure Sensor
STOP
E30
Unrecognized Version of Parameter Storage
STOP
E31
Un-repairable Parameter Storage
STOP
E32
Corrupt Parameter Storage
STOP
E33
Excessive Parameter Storage Size
STOP
E34
Unable to Queue Data to Parameter Storage
REBOOT
E35
Unable to write to parameter storage
STOP
E36
Parameter Out of Range
STOP
E37
Corrupt Real-Time Clock Value
CONTINUE
E38
Real-Time Clock not Ticking
CONTINUE
E39
Unable to Queue Data to User Interface
REBOOT
E40
Invalid Built-in Self Test Call
Not Used
E41
Over Pressure Condition
STOP
E42
Operating System not Responding to Software
REBOOT
E43
Insufficient Voltage for Audible Alarm
STOP
E44
12V Reference Out of Range
STOP
E45
5V Reference Out of Range
STOP
E46
Bulk Voltage Out of Range
STOP
E47
-15V Reference Out of Range
STOP
E48
Railed Flow Sensor
STOP
E49
Blower Pressure Sensor Failure
STOP
E50
Flash Memory Failure
Not Used
E51
Unrecognized Main PCA
STOP
E52
Blower Failure
STOP
E53
Blower Speed Out of Tolerance
STOP