displays if the access level is set to Provider mode. The monitoring screen also displays the actual
measured pressure and the Flex icon if Flex is enabled.
When in the Standby or Monitoring screens, you can modify the humidifier setting by pressing and
holding the Heat button until the following screen appears:
You can increase or decrease the humidifier setting from 1 to 5 in increments of 1. The setting
changes immediately as you adjust it.
P
ROVIDER
M
ODE
There are two ways to access Provider mode:
1. When the access level parameter is set to 1, the BiPAP Synchrony will be in Provider
mode when power is supplied to the unit. The only way to exit this mode is to change the
access level parameter to 0.
2. Simultaneously hold down the right user button and the alarm silence button for approxi-
mately 2 seconds while supplying power to the unit until the device beeps. Pressing the
alarm silence button at any time will exit you from the Provider mode.
S
ETTING
UP
THE
B
I
PAP S
YNCHRONY
1. Enter the Provider Mode. The first screen is the Mode setting screen. To change the
selection, press the Heat or Ramp button until the correct setting appears. Available
modes are PC, CPAP, S, S/T, and T. Press the right user button once the desired setting
is reached.
NOTE
In order to scroll through the different menus, you must use the left and right user
buttons along with the ramp and heat buttons.
S/T
PC