A maximum IPAP time of 3.0 seconds acts as a safety mechanism to limit the time spent at the IPAP
level during spontaneous breathing in the S and S/T modes. Once the time limit is reached, the unit
automatically cycles off IPAP to the EPAP level.
As flow begins to decrease during IPAP, a flow reversal can occur due to a large leak around the mask
or because the patient’s mouth is open. When the BiPAP Synchrony unit senses this flow reversal, the
unit automatically cycles to the EPAP level.
S
YMBOLS
FOR
D
EVICE
AND
P
OWER
S
UPPLY
S
YMBOLS
AND
M
EANINGS
Attention, consult accompanying documents
DC Power
Pressure On/Off
Type BF Applied Part
Class II (Double Insulated)
European CE Declaration of Conformity
Canadian/US Certification