How to use Elisée 350
NOT013281-8
Manual breath (not available in CPAP mode)
Pressing the Manual breath button immediately triggers an additional respiratory cycle.
Measure RC (Invasive ACV mode only)
WARNING
Pressing this button will alter the respiratory cycle while the measurements are being
taken. Patient respiration may be impeded.
This function allows you to measure the patient’s physiological respiratory characteristics:
resistance, compliance, and intrinsic pressure.
The ventilator delivers a square flow cycle with a 2.0 second end-inspiratory pause and a 0.5
second end-expiratory pause.
The formulae used to calculate the resistance and compliance are:
•
R = (Pmax – Pplat) / max
•
Cstat = V
TI
/ (Pplat – PEEPtot)
where:
•
PEEPtot = PEEPi + PEEP (before the end-expiratory pause).
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
cm
H2O
M
od
es
Pa
ra
m
<PE
<PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Freeze-> Touch curves
3/4
Manual
Manual
breath
breath
Measure
Measure
RC
RC
P
0,1
0,1
1. Press screen 3/4
.
2. Press
to
trigger an additional respiratory cycle.
3/4
Manual
Manual
breath
breath
°
V
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
c
m
H2O
M
od
es
Pa
ra
m
<
PE
<
PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Freeze-> Touch curves
3/4
Manual
Manual
breath
breath
Measure
Measure
RC
RC
P
0,1
0,1
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
c
m
H2O
M
od
es
Pa
ra
m
<
PE
<
PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
RC calculation !
R=11,5 cmH2O/l/s
Cstat. = 30 ml/cmH2O
PEEPi = 0,1 cmH2O
1. Press screen 3/4
.
2. Press the
button.
The ventilator performs the calculations and then displays the
results at the bottom of the screen. The results are displayed for
five minutes.
3/4
Measure
Measure
RC
RC
Press to return to the curves screen
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...