Technical specifications
NOT013281-8
Antibacterial filter
Type of filter: Intersurgical Filta-Guard or equivalent:
•
Filter: anti-bacterial / antiviral
•
Connector: 22 female and 22 male / 15 female
•
Bacterial and viral retention: > 99.999%
•
Resistance @ 30 L/min: 1.1 cm H
2
O
•
Resistance @ 60 L/min: 2.2 cm H
2
O
•
Compliance: 0.2 mL/cm H
2
O
•
Compressible volume: 66 mL
•
Internal volume: 200 mL
•
Duration of use: 24 hours.
Humidifier
The humidifier used must meet the following requirements:
•
Maximum operating pressure: > 80 cm H
2
O
•
Max. flow rate: 180 L/min
•
Pressure drop at 180 L/min: < 3 cm H
2
O
•
With spontaneous breathing: < 3 cm H
2
O
•
Air loss at maximum pressure: < 20 mL/min
•
Average compliance: 0.3 to 0.5 mL/cm H
2
O
•
Liquid flow rate: 10 to 25 mg/L.
Nebuliser
Type: Intersurgical Cirrus or equivalent.
Maximum A-weighted sound pressure level: 48.1 dBA
Any brand of nebuliser may be used, provided it is able to function at a maximum pressure of
200 kPa and a maximum flow rate of 20 L/min.
Dust filter
ResMed-specific equipment.
Note:
The manufacturer accepts no liability for damage to the device caused by the use of a
filter which does not comply with the recommendations.
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...