6.2 Alarm types
Type 1: Ventilation monitoring alarms
If two alarms with different priority levels are triggered at the same time, the higher priority alarm
takes precedence over the lower priority alarm. If two alarms with the same priority level are
triggered at the same time, refer to the table under the heading “Relative alarm priorities” on
page 73.
Note:
The ‘Trigger threshold [can be stopped?]’ column in the following table indicates for each
alarm whether it can be stopped permanently (YES), stopped temporarily (2 min) or not stopped
at all (NO).
Symbol
Message
Alarm
Priority
Trigger
threshold [can
be stopped?]
Time to trigger
MINI Vt
Minimum Expired Tidal
Volume
(double circuit)
High
Alarm setting
[2 min]
5 consecutive cycles above
alarm threshold
MAXI Vt
Maximum Expired Tidal
Volume
(double circuit)
High
Alarm setting
[2 min]
5 consecutive cycles above
alarm threshold
MINI Vt
Minimum Inspired Tidal
Volume
(single circuit)
High
Alarm setting
[2 min]
5 consecutive cycles above
alarm threshold
MAXI Vt
Maximum Inspired Tidal
Volume
(single circuit)
High
Alarm setting
[2 min]
5 consecutive cycles above
alarm threshold
MINI V/min
Minimum Expired Minute
Volume
(double circuit)
High
Alarm setting
[2 min]
Immediately
MAXI V/min
Maximum Expired Minute
Volume
(double circuit)
Low
Alarm setting
[2 min]
Immediately
MINI V/min
Minimum Inspired Minute
Volume
(single circuit)
High
Alarm setting
[2 min]
60 s
MAXI V/min
Maximum Inspired Minute
Volume
(single circuit)
Low
Alarm setting
[2 min]
60 s
MAXI P
INSPI*
Maximum Inspiratory
Pressure
High
Alarm setting
[2 min]
3 consecutive cycles above
alarm threshold
MAXI P EXPI*
Maximum Expiratory Pressure
High
10 mbar + PEEP
[2 min]
300 ms above alarm threshold
LP – Low
flow?
Minimum Pressure (volume-
controlled modes)
Double circuit
High
3 mbar + leak
< 80%
[2 min]
2 complete cycles below the
threshold or maximum of 17 s
CIRCUIT
OPEN
Minimum Pressure (volume-
controlled modes)
Double circuit
High
3 mbar + leak
≥
80% [2 min]
2 complete cycles below the
threshold or maximum of 17 s
CIRCUIT
OPEN
Minimum Pressure (volume-
controlled modes)
Single circuit
High
Pi alarm
threshold
[2 min]
2 complete cycles below the
threshold or maximum of 17 s
CIRCUIT
OPEN
Minimum Pressure (pressure-
controlled modes)
High
(Pi or PS) – 2 mbar
at the end of
inspiration
[2 min]
2 complete cycles below the
threshold or maximum of 17 s
CIRCUIT
OPEN
Minimum pressure (pressure-
controlled modes, CPAP)
High
(CPAP+1)/3 during
inspiration [2 min]
2 complete cycles below the
threshold or maximum of 17 s
LOW PEEP
Minimum Expiratory Pressure
(double circuit)
Medium
Alarm setting
[2 min]
30 cycles below alarm
threshold
MAXI F
Maximum Respiratory Rate
Low
Alarm setting
[2 min]
4 consecutive cycles above
alarm threshold
L
L
L
Δ
L
L
L
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...