Alarms and troubleshooting
NOT013281-8
Tests which apply to both single and double circuit alarms
1. Connect a patient circuit (single or double) fitted with a Maquet test lung to the ventilator and
perform the manual test.
2. Set the mode to ACV, Adult.
3. Set the ventilation parameters as follows:
4. Start ventilation and perform the following actions:
5. Connect a high-pressure oxygen supply and a 2 litre test lung to the ventilator, then set the
following values:
6. Set the V
TE
, Ve, and Pmax alarm thresholds to their respective minimum and maximum
values.
7. Start ventilation and perform the following actions:
8. Disconnect the oxygen and change the ventilation mode to PS.
9. Set the ventilation parameters as follows:
V
T
= 450 mL
max = 28 L/min
PEEP = 0 cm H
2
O Trigger
=
NO
F = 15 bpm
FiO
2
= 21%
Alarm
Action
Visual alert
Audible
signal
Time the
alarm takes
to trigger
CIRCUIT OPEN
Disconnect the patient circuit expiratory limb
Flashing
Sounding
2 cycles
Reconnect the patient circuit expiratory limb
Steady
Off
Press the Alarm Silence button
Off
Off
MAXI F
Adjust the upper alarm threshold to F = 10 bpm
Flashing
Sounding
4 cycles
Adjust the upper alarm threshold to F = 35 bpm
Steady
Off
Press the Alarm Silence button
Off
Off
MAXI P INSPI
Adjust the upper alarm threshold to Pmax = 10 cm H
2
O
Flashing
Sounding
3 cycles
Adjust the upper alarm threshold to Pmax = 60 cm H
2
O
Steady
Off
Press the Alarm Silence button
Off
Off
V
T
= 2000 mL
max = 53 L/min
PEEP = 5 cm H
2
O
Trigger = 2
F = 10 bpm
FiO
2
= 21%
Alarm
Action
Visual alert
Audible
signal
Time the alarm
takes to trigger
MINI FIO
2
Set FiO
2
to 100%
Flashing
Sounding
2 minutes
Set FiO
2
to 21%
Steady
Off
Press the Alarm Silence button
Off
Off
PS = 15 cm H
2
O
PEEP = 5 cm H
2
O
Ti = 1.2 s
Tapnea = 15 s
F = 15 bpm
Pi = 15 cm H
2
O
Trigger = 3
FiO
2
= 21%
°
V
°
V
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...