Technical specifications
NOT013281-8
Ventilator with double circuit (Ø 22 mm), product code CIR009727 & accessories (filter,
humidifier, etc.)
•
Device volume: 1700 mL
•
Device compliance: 1.7 mL/cm H
2
O
•
Inspiratory resistance @ 60 L/min: 4.3 cm H
2
O
•
Expiratory resistance @ 60 L/min: 3.0 cm H
2
O.
High-pressure oxygen
•
High-pressure oxygen port: 240–700 kPa, 150 L/min maximum
•
In (A)CV mode, with V
T
= 500 mL, PEEP = 7 cm H
2
O, F = 15 bpm, I:E = 1:2.0 and
FiO
2
= 100%: O
2
consumption = 18 L/min ± 20%.
Operating, storage and transport
Normal operating conditions
•
Ambient temperature when the device is running on battery power: -10°C to +40°C (14 to
104°F)
•
Ambient temperature required for recharging the batteries during operation: +5°C to +40°C
(41 to 104°F)
•
Ambient relative humidity: 10% – 95%
•
Atmospheric pressure: 600 to 1100 hPa.
CAUTION
The performance of the device is not guaranteed in operating conditions other than those
described above.
Storage conditions
•
Storage temperature: -10°C to +50°C (14 to 122°F)
•
Ambient relative humidity: 10–90%
•
Atmospheric pressure: 500 to 1100 hPa
•
The device is fragile. Keep dry and store in its operating position or in its bag.
CAUTION
We recommend that you store the device in its packaging or in its bag. If the device is stored
outside of its packaging, it must be placed flat on its base.
Conditions for transport when not delivering ventilation
When returning to the service centre, the ventilator and its accessories must be transported
inside the ResMed packaging. They must be transported in the following conditions:
•
Storage temperature: -10°C to +50°C (14 to 122°F)
•
Relative humidity: 10–90%
•
Atmospheric pressure: 500 to 1100 hPa.
CAUTION
The device is fragile.
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...