How to use Elisée 350
NOT013281-8
1 Checking the power supplies
Three LEDs on the front of the device indicate the charging status of the power supplies, whether
the ventilator is on standby or delivering ventilation.
Notes:
•
When operating the device solely on battery power, ensure that the battery is sufficiently
charged (green LED flashing slowly). If not, plug the ventilator into the mains power supply
and switch on (if necessary).
•
The battery charging status is displayed one minute after the ventilator is switched on. Each
dash
indicates a charge of around 25%.
•
In the event of power failure, the event log and all of the settings will be saved, no matter
how long it takes to restore power.
Conditions for recharging the internal and external batteries
Note:
You should perform a complete charge/discharge cycle for internal and external batteries
before use. Internal and external batteries reach optimal performance after three full charge/
discharge cycles.
If an external DC power supply (12–28 V DC / 15 A max) is in use, it must have enough charge to
supply power for at least one hour.
Power supply status display
The following table lists the power supply symbols as well as the symbols that show whether the
(internal/external) battery is charging or discharging.
Power supply
Symbol and LED status
Meaning
Mains power pack or
external power
supplies
(green LED, not flashing)
Mains power pack or external DC supply
present
(LED off)
Neither mains power pack nor external
DC supply present
Internal battery or
external battery pack
(green LED, not flashing)
Battery charged
(green LED flashing slowly)
Battery discharging
(orange LED flashing slowly)
Battery discharging (battery low)
(red LED flashing slowly)
Battery flat
(orange LED flashing quickly)
Battery charging interrupted (battery
temperature too high or too low)
(green LED flashing quickly)
Battery charging
Symbol
Meaning
Internal battery present
External battery present
External power supply present
Mains supply present
E
X
T
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...