WARNING
In P(A)CV, the flow is adjusted so that a constant pressure is maintained. Consequently, the
volume delivered may vary depending on compliance and resistance. It is therefore
important to monitor the inspiratory and, in particular, the expiratory volumes through the
min. and max. V
TI
and V
TE
alarms.
The adjustable parameters are as follows:
PCV mode:
PACV mode:
Adjustable parameters (PVC/PACV)
Pi
F
Ti
PEEP
Rise time
Inspiratory pressure and flow triggers (PACV only)
Recruitment parameters (period, size, pressure)
FiO
2
P
V
t (s)
t (s)
t (s)
CC
CC
CC
CC
Ti
Ti
Ti
Ti
Vol.
CC: Controlled Cycle
Ti: Set Ti
Set pressure (Pi)
t (s)
t (s)
t (s)
P
V
Ti
Ti
CC
ACC
CC
CC
t (s)
t (s)
t (s)
Vol.
CC: Controlled Cycle
ACC: Assisted Controlled Cycle
Ti: Set Ti
Set pressure
Patient effort
t (s)
t (s)
t (s)