NOT013281-8
10. Start ventilation and perform the following actions:
Tests for double circuit-only alarms
1. Connect a double patient circuit fitted with a Maquet test lung to the ventilator and perform
the manual test.
2. Set the mode to ACV, Adult.
3. Set the ventilation parameters as follows:
4. Start ventilation and perform the following actions:
5. Set the mode to PACV, Adult and perform the following actions:
Alarm
Action
Visual alert
Audible
signal
Time the alarm
takes to trigger
PATIENT APNEA
Trigger 4 inspiratory cycles
Steady
Off
Tapnea
Wait without triggering further inspiratory phases
Flashing
Sounding
Press the Alarm Silence button
Off
Off
O
2
FAULT
Set the FiO
2
to 90%
Flashing
Sounding
2 minutes
Set the FiO
2
to 21%
Steady
Off
Press the Alarm Silence button
Off
Off
V
T
= 500 mL
max = 28 L/min
PEEP = 0 cm H
2
O
Trigger =
NO
F = 15 bpm
FiO
2
= 21%
Alarm
Action
Visual alert
Audible
signal
Time the alarm
takes to trigger
MINI Vt
Adjust the upper alarm threshold to V
TE
= 2000 mL and the
lower alarm threshold to V
TE
= 800 mL
Flashing
Sounding
5 cycles
Adjust the lower alarm threshold to V
TE
= 150 mL
Steady
Off
Press the Alarm Silence button
Off
Off
MAXI Vt
Adjust the upper alarm threshold to V
TE
= 250 mL
Flashing
Sounding
5 cycles
Adjust the upper alarm threshold to V
TE
= 2000 mL
Steady
Off
Press the Alarm Silence button
Off
Off
MAXI V/min
Adjust the lower alarm threshold to
e = 0.2 L/min and the
upper threshold to
e = 6.5 L/min
Flashing
Sounding
Immediately
Adjust the upper alarm threshold to
e = 30.0 L/min
Steady
Off
Press the Alarm Silence button
Off
Off
MINI V/min
Adjust the lower alarm threshold to
e = 11.0 L/min
Steady
Off
Immediately
Adjust the lower alarm threshold to
e = 0.2 L/min
Off
Off
Press the Alarm Silence button
Off
Off
Alarm
Action
Visual alert
Audible
signal
Time the alarm
takes to trigger
MAXI LEAKS
Adjust the upper alarm threshold to Leak = 20% then create a
slight leak in the circuit (e.g. by slightly loosening the
expiratory limb of the circuit)
Flashing Sounding
6 cycles
Adjust the upper alarm threshold to Leak = 80% and stop the
leak
Steady
Off
Press the Alarm Silence button
Off
Off
°
V
°
V
°
V
°
V
°
V
°
V
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...