Ventilation
NOT013281-8
Cycling occurs at the end of the inspiratory time (Ti) set by the user. For volume modes only, Ti
is set either directly by means of the Ti setting, or indirectly by means of the I:E or max settings,
see “Selecting target parameters in ACV mode (during ventilation and while ventilation is
stopped)” on page 63.
The adjustable parameters are as follows:
Note:
In (A)CV mode, the Elisée 350 delivers flow-controlled breaths. The airway pressure and
the alveolar pressure depend on the resistance and compliance of the patient’s respiratory
system.
WARNING
To avoid the risk of barotrauma, it is important to set the High Pressure alarm (Pmax) and
to monitor the measured values, in particular the plateau pressure if a plateau time was set.
(Assisted) Pressure-Controlled Ventilation (P(A)CV)
In P(A)CV mode, cycles are pressure-controlled. The flow is adjusted so that the set pressure is
maintained during a set inspiratory time.
The target Pi is added to the PEEP, so Total inspiratory pressure = Pi + PEEP. Controlled cycles
are delivered according to the respiratory rate set by the user. In PACV mode, assisted controlled
cycles can be triggered by the patient over and above the set respiratory rate.
The tidal volume and the minute volume are based on the resistance and compliance of the
patient’s respiratory system.
Adjustable parameters (CV/ACV modes)
V
T
F
Ti, max. flow (
max) or I:E
Flow shape
Tplateau
PEEP
Inspiratory pressure and flow triggers (ACV only)
Recruitment parameters (period, size, pressure)
FiO
2
°
V
°
V
P
V
Vol.
t (s)
t(s)
t(s)
CC
CC
ACC
Ti
Ti
Ti
CC: Controlled Cycle
ACC: Assisted Controlled Cycle
Ti: Set Ti
Patient effort
t (s)
t (s)
t (s)
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...