Setting up
NOT013281-8
Active humidification system
A humidification system makes it possible to increase the amount of moisture in the air delivered
to the patient. Set up the heater base and connect it to a power supply in accordance with the
manufacturer’s instructions.
Note:
If a heated circuit is being used with a humidification system, connect it to a power
supply, then wait 15 minutes before performing a manual test so that the temperature and
humidity can stabilise. This will ensure accurate measurement of the V
TE
(when using a double
circuit).
WARNING
Set up the humidifier in accordance with the manufacturer’s instructions. Always ensure
that the ventilator and expiratory valve are placed above the humidifier and water traps.
Note:
The water traps must be emptied periodically.
For further details, refer to the user guide for this accessory.
Standard pneumatic nebuliser
The driving gas is pure oxygen with a maximum pressure of 200 kPa and a flow rate of 20 L/min.
Follow the manufacturer’s instructions for assembling and connecting the nebuliser.
CAUTION
Do not insert a filter between the patient and the nebuliser.
Remote alarm
For further details, refer to the user guide for this accessory.
Water trap (optional)
Heated humidifier
Antibacterial filter
3
1
2
1. Connect the high-pressure oxygen hose.
2. Connect the nebuliser tubing.
3. Connect the nebuliser between the end of the
patient circuit and the patient interface.
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...