53/102
How to use Elisée 350
NOT013281-8
3 Setting alarm thresholds
Setting the
Pi alarm threshold – Special case
CAUTION
If you change any of the alarm settings, see Section 6.5 on page 74 for instructions on
testing the alarms.
This setting is used:
•
Usually for ventilation in volume-based modes
•
With a single circuit. When using a single circuit, this is the only way of detecting when the
patient circuit is disconnected.
Pi represents the minimum pressure threshold to be maintained during inspiration. If the
pressure falls below this threshold, the Low Pressure alarm is triggered.
Pi = Pmax – PEEP, where Pmax is the maximum inspiratory pressure measurement.
Example of how to set the
Pi alarm threshold, with PEEP at 5 cm H
2
O and the
Pi value at
7 cm H
2
O:
0
10
20
30
40
50
60
70
80
90
<PE
0
H
2
O
cm
S
T
A
R
T
S
T
A
R
T
V
E
N
T
I
L
V
E
N
T
I
L
Pa
ra
m
m
ACV
ACV
info
bpm
ml
cmH2O
%
35
35
1000
1000
15,0
15,0
3,0
3,0
300
300
40
40
30
30
F
Pmax
Pmax
Leak
Leak
Vte
Vte
V
E
0
0
0
- -- -
0,0
M
od
es
s
┘
HP
0
10
20
30
40
50
60
70
80
90
<PE
┘
HP
00
H
2
O
cm
info
Parameters settings
ACV
ACV
Ventil.
Ventil.
stop
stop
Ventil.
Ventil.
stop
stop
Ventil.
Ventil.
stop
stop
PEEP
PEEP
0
cm
H
2
O
FiO
FiO
2
60
%
V
T
I:E
I:E
500
ml
F
15
1:2.0
bpm
1/3
S
T
A
R
T
S
T
A
R
T
V
E
N
T
I
L
V
E
N
T
I
L
M
od
es
OK
OK
Alarms setting
Alarms setting
ACV
ACV
0
10
20
30
40
50
60
70
80
90
<PE
00
H
2
O
cm
info
bpm
ml
l/min
l/min
cmH2O
%
35
35
1000
1000
15,0
15,0
3,0
3,0
300
300
30
30
F
Pmax
Pmax
Leak
Leak
Vte
Vte
V
E
0
0
0
- -- -
0,0
50
50
Cancel
Cancel
┘
HP
1. Press .
2. Press the value to be modified (here, Pmax = 50).
3. Press
or
to set the value (a flashing message appears at
the top of the screen when the upper or lower limit is
reached), then press
to confirm.
Note:
It is possible to change several alarm thresholds in
succession and confirm them all at once.
Now you can:
•
Set the ventilation mode again by pressing
•
Adjust the parameters again by pressing
.
OK
OK
M
od
es
P
Paa
rraa
m
m
Δ
Δ
Δ
Δ
Δ
Cancel
Cancel
0
10
20
30
40
50
60
70
80
90
<
PE
┘
HP
00
H
2
O
cm
info
Para
m
eters setting
ACV
ACV
Ventil.
Ventil.
stop
stop
Ventil.
Ventil.
stop
stop
FiO
FiO
2
60
%
V
T
PEEP
PEEP
500
m
l
5
c
m
H
2
O
Vmax
Vmax
F
OK
OK
60
m
in
15
bp
m
l/
Alar
m
s setting
ACV
ACV
0
10
20
30
40
50
60
70
80
90
<
PE
00
H
2
O
cm
M
od
es
s
info
S
T
A
R
T
S
T
A
R
T
V
E
N
T
I
L
V
E
N
T
I
L
Pa
ra
m
m
40
40
bp
m
m
l
c
m
H2O
35
35
1000
1000
15,0
15,0
3,0
3,0
300
300
7
F
Pmax
Pmax
Vti
Vti
V
i
0
0
l/
m
in
0,0
0,0
0
∆
Pi
Pi
c
m
H2O
0
┘
HP
1/3
1. On the parameter settings screen, set the target PEEP to
5 cm H
2
O.
2. On the alarms screen, set
Pi to 7 cm H
2
O. To avoid triggering
the Low Pressure alarm, the pressure during inspiration must
always exceed the threshold of 7 cm H
2
O above PEEP.
Δ
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...