3 Reassembly
Replacing the dust filter
Replacing the oxygen sensor
Note:
The replacement frequency for the oxygen sensor varies from 6 to 12 months. This
frequency is given as an indication only and varies depending on the duration of ventilation, the
oxygen concentration used and the ambient temperature. When the sensor expires, the
ventilator will tell you that it is time to replace it (see “3 Manual test” on page 48).
58
58
1
2
58
NE
W CLAPPER
1. After autoclaving or disinfecting the expiratory valve,
place a clapper in the holder so that the side with a
number on it is visible, as in the figure opposite.
2. Line up the three notches with the locking nibs and
push firmly.
Notches
Widest
notch
4
3
3. Lock the support together by turning the clapper
holder clockwise as far as you can.
4. Replace the expiratory valve by matching up all three
pressure connectors with the corresponding holes on
the valve.
Pressure
connector
1. Use tweezers to pull the filter out of its
compartment.
2. Insert the new filter into the compartment and
push it in until flush with the casing.