2 Checking the patient circuit
Before connecting the patient to the ventilator, check the whole of the patient circuit to eliminate
any risk of leaks or defective parts.
Check that:
•
The expiratory circuit support matches the circuit type: single or double
•
The patient circuit and accessories are properly connected to the expiratory circuit support.
Note:
Perform a manual test to ensure the ventilator has correctly identified the type of patient
circuit installed.
3 Manual test
This test must always be performed before the ventilator is programmed for a new patient and
whenever the circuit configuration is changed.
WARNINGS
•
The ventilator must not be connected either to a patient or to an oxygen supply during
this test.
•
If the test is unsuccessful, check the circuit configuration and repeat the test. Make sure
you complete every step. If the test is again unsuccessful, stop using the ventilator and
contact a technician.
•
Make sure that when you perform this test, all the accessories that will be used during
ventilation are connected. Repeat the test if an accessory is added or removed.
This quick test (around 1 minute) consists of four stages:
•
Measurement of the sensor offsets
•
Checking for internal leaks in the ventilator
•
Measurement of the inspiratory resistance of the circuit
•
Measurement of the expiratory resistance and calibrating the expiratory flow (if a double
circuit is connected), the circuit compliance and the O
2
sensor gain.
RESTART
RESTART
TEST
TEST
Last test
Ctot = 0,7 ml/cmH2O
Kind of circuit : DOUBLE
Circuit test : PASS
le 18/06/10 at 14 : 50 : 35
Ri = 0.4 cmH2O (30 l/min)
Ri = 0.7 cmH2O (60 l/min)
Re = 1.4 cmH2O (30 l/min)
Re = 2.9 cmH2O (60 l/min)
Back
CONTINUE
CONTINUE
Unplug patient circuit !
Unplug oxygen
Cancel
When done, press:
PRESET
PRESET
PREVIOUS SETTING
PREVIOUS SETTING
ACV
ACV
SETTING
SETTING
TEST
TEST
Las test the 18/06/10
Double circuit : PASS
NEW
NEW
SETTING
SETTING
1. On the main screen, press the “Test” button.
The results of the last test are displayed.
2. Press the “Restart” button.
3. Follow the instructions displayed (disconnect the
oxygen supply, the patient circuit and its
accessories), then press “Continue”. The ventilator
measures the sensor offsets.
Return to main
screen
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...