NOT013281-8
6.6 Troubleshooting
WARNING
If, after trying these solutions, the activated alarm persists, turn the device off and contact
your technician.
Alarms that cannot be stopped
Alarms that can be stopped temporarily (muted for 120 seconds)
Symbol
Message
Priority
Cause
Solution
INT BATT EMPTY
High
Internal battery charge less than
5%
Change the power supply
MINI FIO
2
High
Supplied FiO
2
too low
Check that an oxygen supply is connected and
press the
button (see page 65)
MAXI FIO
2
High
Supplied FiO
2
too high
Check that an oxygen supply is connected and
press the
button (see page 65)
NiMH BATTERY
Low
An unknown external battery
pack has been connected
Connect a compatible external battery pack
(Li-Ion type)
O
2
100%
2 min
O
2
100%
2 min
Symbol
Message
Priority
Cause
Solution
O
2
FAULT
High
HP O
2
supply missing
Connect the high-pressure O
2
tube
PATIENT APNEA
Medium
No spontaneous cycles were
inititated during Tapnea starting
from the end of the previous
expiratory phase
Check for leaks from the mask or tubing.
If necessary, adjust the apnea time.
For the double circuit only, adjust the trigger
value
EXT BATT EMPTY
Medium
External battery charge less
than 5%
Connect the device to mains supply or change
the external battery pack
EXT BATT LOW
Low
External battery charge less
than 15%
Connect the device to mains supply or change
the external battery pack
INT BATT LOW
Medium
Internal battery charge less than
15%
Change the power supply
CIRCUIT OPEN
High
Patient circuit is poorly
connected
Reconnect the patient circuit
Patient circuit is faulty
Change the patient circuit
TEST FAILED
High
Manual test failed
Repeat the manual test
MAXI F
Low
Respiratory rates exceed the
upper threshold of the
Respiratory Rate alarm
Check that Respiratory Rate alarm and
inspiratory trigger settings are correct
MAXI LEAKS
High
Leaks exceed the upper
threshold of the High Leak
alarm
Check that the High Leak alarm threshold
setting is correct and check the integrity of
the patient circuit
WRONG CIRCUIT
High
Type of patient circuit used is
different from the one
determined by the test
Check the circuit and repeat the test
LOW PEEP
Medium
The PEEP delivered to patient is
lower than the target PEEP
Check the integrity of the patient circuit
MAXI P EXPI
High
The expiratory pressure is too
high compared with the target
PEEP
Check the expiratory valve control and check
the patient circuit is not blocked
L
L
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...