2.4 Power supplies
The Elisée 350 can be used with several different power supplies. The device automatically
selects the power supply to be used, according to the following hierarchy:
1. A mains power pack installed in the compartment under the device.
2. An external power supply connected to the Elisée external DC power socket, which could be:
•
A mains power pack located ouside the ventilator
•
An external power supply (12
–
28 V DC) via the cigarette lighter.
3. An external battery pack installed in the compartment under the device.
4. The internal battery.
CAUTION
Only power supplies or power cords designed and distributed by ResMed should be
connected to the ventilator.
Note:
In the event of a loss of mains power, an audible and visual alarm is activated. It can be
stopped by pressing the Alarm Silence button.
–
Ventilator serial number label. E.g. EL2H1101001:
Product
code
Year of
manufacture
Month of
manufacture
Manufacture
No.
EL2H
11
01
001
Manufacturer plate
Compartment for mains power pack or external battery
pack
Product code (PACXXX)
Removable dust filter
Speaker
4
1
2
3
Internal battery
Mains power pack (grey label and
two connectors)
External battery
pack (blue label,
no connectors)
Mains power pack with external power cord (in this case, the
pack is treated as an external power supply)
External power
supply (via vehicle
cigarette lighter)
External power supplies
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...