NOT013281-8
Respiratory mechanics parameters
On the curves screen, two screens display the ventilatory mechanics parameters, which can be
accessed using the
(screen 2/4) button and the
(screen 3/4) button.
Plateau time
Pressing the End-inspiratory pause button inserts a plateau time at the end of the inspiratory
cycle, once the minimum tidal volume has been reached. Hold your finger on the button for the
desired length of the plateau time (max. 5 s). The ventilator will keep the inspiratory and expiratory
valves closed, and no air flow will be delivered.
End-expiratory pause
Pressing the End-expiratory pause button inserts a pause at the end of the expiratory cycle.
During the pause, the ventilator keeps the inspiratory and expiratory valves closed.
2/4
3/4
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
cm
H2O
M
od
es
Pa
ra
m
<PE
<PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Freeze -> Touch curves
2/4
Nebul.
Nebul.
I
E
Pause
Pause
Pause
Pause
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
26
cm
H2O
<PE
<PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Frozen curves !
Back
Back
O
K
Ppeak = 26,4 cmH2O
Pplat = 22,4 cmH20
R = 9,3 cmH2O/l/s
1. Press screen 2/4
.
2. Then press
during the inspiratory phase to insert an end-
inspiratory pause. The maximum pause time is
5 seconds
,
after which expiration is automatically triggered.
The screen displays the plateau pressure (Pplat), maximum
pressure (Ppeak) and resistance (R) values.
3. Press “OK” to go to the frozen curves screen (see “Freezing
to return to the curves screen.
Note:
Wait for the real-time curves to be displayed at the left-
hand side of the screen before pressing the Pause button.
Otherwise, the frozen curves will not display correctly.
2/4
I
Pause
Pause
Back
Back
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
cm
H2O
M
od
es
Pa
ra
m
<PE
<PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Freeze -> Touch curves
2/4
Nebul.
Nebul.
I
E
Pause
Pause
Pause
Pause
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
27
cm
H2O
<PE
<PE
P
14
0
28
0
>
>
v
O
2
100%
2
min
Frozen curves !
Back
Back
O
K
PEEPtot= 0,5 cm H2O
PEEPi = 0,1 cmH2O
1. Press screen 2/4
.
2. Then press
during the expiratory phase to insert an end-
expiratory pause. The maximum pause time is
12 seconds
,
after which inspiration is automatically triggered.
The screen displays the total positive expiratory pressure
(PEEPtot) and the intrinsic positive expiratory pressure (PEEPi).
3. Press “OK” to go to the frozen curves screen (see “Freezing
to return to the curves screen.
Note:
Wait for the real-time curves to be displayed at the left-
hand side of the screen before pressing the Pause button.
Otherwise, the frozen curves will not display correctly.
2/4
E
Pause
Pause
Back
Back
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...